GMAp Space
GMAp's single physical space, in the Parobé Building. Comprises experimental area, computational area, faculty room, classroom, and project room. Formed in 1994 from the union of three earlier laboratories (LVDE, LMM, LRM) — the historical names are on the History page.
- Total area 500 m²
- Internal zones 5
- Location Parobé Building, DEMEC / School of Engineering / UFRGS
Internal zones
-
Experimental area
Space dedicated to the group's physical testing and experimental characterization, with instrumentation benches for mechanical, dynamic and acoustic testing. It hosts the measurement, prototyping and experimental validation work that supports GMAp's research lines.
-
Computational area
The group's computational workspace, with workstations for numerical modeling and simulation — finite element method, structural optimization and mechanical systems analysis. This is where the models that complement the experimental work are developed.
-
Faculty room
Workspace for GMAp's professors and researchers, dedicated to student supervision, scientific writing and the coordination of the group's projects.
-
Classroom
Room for graduate courses, seminars and thesis defenses, integrating teaching with the research activities carried out in the space.
-
Project room
Shared workspace for students and research teams, where master's and doctoral students and undergraduate research fellows develop their projects.
Equipment
- Universal testing machine
Shimadzu Autograph AG-Xplus
- Capacity
- 250 kN
- Temperature
- room to 300 °C
- Extensometer
- clip-gauge; SES-1000 (rubber, up to 1,000 mm); TRViewX (optical)
- Tests
- tension, compression, bending
- Universal testing machine
Timegroup WDW-100E
- Capacity
- 100 kN
- Extensometer
- clip-gauge
- Tests
- tension, compression, bending, shear
- Universal testing machine
EMIC DL2000
- Capacity
- 20 kN frame, 5 kN load cell
- Temperature
- room to 150 °C
- Extensometer
- clip-gauge
- Tests
- tension, compression, bending, Arcan
- Dynamic testing
MTS Acumen 3
- Capacity
- 2 kN static, 3 kN dynamic
- Frequency
- up to 100 Hz
- Stroke
- 70 mm
- Temperature
- chamber, −150 °C to 350 °C; bath, 5 °C to 40 °C above ambient
- Medium
- Bionix EnviroBath saline bath, 10 L
- Tests
- static, fatigue, DMA, DTMA
- Dynamic testing
TA ElectroForce 3200
- Capacity
- 225 N, 5.6 N·m (axial-torsional)
- Frequency
- 0.00001 Hz to 300 Hz
- Stroke
- ±6.5 mm; extender up to 150 mm
- Temperature
- bath, room to 45 °C
- Medium
- saline bath
- Tests
- static, fatigue, DMA up to 200 Hz
- Electrodynamic shaker
LDS V550
- Frequency
- 5 Hz to 6,300 Hz
- Payload
- up to 25 kg
- Electrodynamic shaker
VEB Thüringer
- Frequency
- 20 Hz to 5,000 Hz
- Payload
- up to 60 kg
- High strain rate
Hopkinson bar
- Strain rate
- 10² to 10⁵ s⁻¹
- Length
- over 6 m
- High-speed camera
Phantom v411
- Full resolution
- 1280 × 800 at 4,200 fps
- Top rate
- 600,000 fps at 128 × 8
- Bit depth
- 12 bits
- Digital image correlation
Optical bench
- Measurement
- 2D and 3D full-field displacements
- Optics
- several lenses and cameras
- Calibration
- certified calibration targets
Also: a strain-gauge and signal-acquisition system; a Charpy impact tester; IPT dynamometric standards for force calibration; in-house numerical simulation software and commercial licences for optimization and simulation.
Laboratory coordination
Testing and registration
- Tensile
- Compression
- Flexural (bending)
- Fatigue
- Impact (Charpy)
- DMA — dynamic mechanical analysis